especificaçoes:

E3610A 30W Power Supply, 8V, 3A or 15V, 2A

http://cp.literature.agilent.com/litweb/pdf/5968-9727EN.pdf

E3610A, E3611A, and E3612A
single-output, dual range
models
These popular 30-watt bench
supplies are designed for general
laboratory use. The constant-voltage,
constant-current output allows
operation as either a voltage or current
source. The changeover occurs
automatically, based on the load.
Each of these models has two ranges,
allowing more current at a lower
voltage. For higher output voltages,
supplies can be connected in series.
These models also feature overload
protection. A continuously acting
constant current circuit protects the
power supply against all overloads
including a direct short placed across
the terminals in a constant voltage
operation.

Specifications
E3610A E3611A E3612A E3614A E3615A E3616A E3617A E3620A E3630A
Features Dual range, 10 turn pots, Constant
Voltage (CV), Constant Current (CC)
modes.
Adjustable overvoltage protection, voltage &
resistance programming, remote sense, rear
outputs, ten turn pots, CV, CC modes. Multiple
supplies can be connected for tracking or higher
power.
Isolated
dual
outputs, 10
turn pots
CV, CL
Tracking, CV,
CL (±20 V)
CV, CF (+6 V)
Number of outputs 1 23
Number of Output
Ranges
2 2 2 1 1 11 1 1
DC Output Rating 8 V, 3 A
15 V, 2 A
20 V, 1.5 A
35 V, 0.85 A
60 V, 0.5 A
120 V, 0.25 A
8 V, 6 A 20 V, 3 A 35 V, 1.7 A 60 V, 1 A 25 V, 1 A
25 V, 1 A
+6 V, 2.5 A
+20 V, 0.5 A
%20 V, 0.5 A
Load and Line
Regulation < 0.01% + 2 mV
Ripple and Noise (20 Hz to 20 MHz)
Normal mode voltage < 200 μVrms, < 2 mVpp < 200 μVrms, < 1 mVpp < 350 μVrms, < 1.5 mVpp
Normal mode current
< 200 μArms, < 1 mVpp < 0.02% +
3 mA
< 0.02% +
1.5 mA
< 0.02% +
1 mA
< 0.02%
+ 0.5
mA

Common mode current Not specified < 1 μArms
Transient Response
Time
< 50 μsec following a change in output current from full load to half load for output to recover within:
10 mV 15 mV
Meter Accuracy ±0.5% + 2 counts at 25 ºC ±5 ºC
Meter Resolution
Voltage 10 mV 100 mV 100 mV 10 mV 10 mV (0%20 V), 100 mV (>20 V) 10 mV
Current 10 mA 10 mA 1 mA 10 mA 10 mA 1 mA 1 mA 1 mA 10 mA
Isolation 240 Vdc